Expanding urolithiasis treatment: comparison of super pulsed thulium laser and holmium:YAG laser for ureteral stone management
European Journal of Medical Research volume 30, Article number: 379 (2025)
Abstract
Objectives
The purpose of this review is to compare the effectiveness of SuperPulsed thulium fiber laser (SP TFL) and holmium:yttrium–aluminum–garnet (Ho:YAG) laser in lithotripsy, with the aim of evaluating the differences between the two in key indicators, such as lithotripsy efficiency and safety, and providing reference for clinical selection of better lithotripsy methods.
Methods
By searching multiple authoritative medical databases (PubMed, Embase, and Web of Science databases) and including the results of relevant clinical studies and laboratory studies, the indexes involving SP TFL and Ho:YAG lasers in the included literature were analyzed.
Results
We found a total of 24 relevant pieces of literature. The laser parameters, such as ablation efficiency, ablation speed, operative time, dust quality, retropulsion, visibility, temperature safety, and stone-free rate, were compared between laboratory studies and clinical outcomes. Preclinical studies have shown that SP TFL has a higher rate of stone ablation, a weaker retropulsion and a lower risk of fiber breakage. The results of clinical studies showed that the two methods were comparable in the ablation rate, laser time and operative time, stone-free rate and complication. SP TFL offered better endoscopic view quality and less retropulsion.
Conclusions
While the Ho:YAG laser remains the primary choice for endoscopic laser lithotripsy, the emergence of SP TFL offers a promising new option for the minimally invasive treatment of urinary calculi. Parameter range, retropulsion effect, laser fiber adaptability, and overall system performance demand comprehensive attention. SP TFL has a relatively short clinical application history, and further research is necessary to fully explore its long-term advantages, clinical significance, and possible limitations.
Introduction
The management of ureteroscopic lithotripsy has evolved significantly with advancements in laser technology. For over two decades, the holmium:yttrium–aluminum–garnet (Ho:YAG) laser has been the gold standard for endoscopic lithotripsy, offering high stone removal rates, minimal tissue damage, and low complication rates [1,2,3]. Despite its proven efficacy, the Ho:YAG laser has limitations, including a maximum frequency of 100 Hz, a multi-mode spatial beam profile that restricts use with fibers smaller than 200 μm, and stone retropulsion during fragmentation [4]. These limitations pose significant challenges, particularly during the management of lower pole calculi or complex cases necessitating maximal endoscopic deflection.
In response to these limitations, thulium fiber laser (TFL) has emerged as a promising alternative. Operating at a wavelength of 1940 nm, TFL aligns with the peak of water absorption and achieves a fourfold higher absorption coefficient. Therefore, when operating at comparable pulse energy parameters, the TFL exhibits dual advantages: a reduced energy threshold required for ablation initiation and significantly increased ablation efficiency. In vitro studies have shown that compared with the Ho:YAG laser, the stone ablation rate of TFL was increased by four times [4,5,6]. In addition, TFL produces a high-quality beam that can be transmitted through smaller fibers (as narrow as 50 μm vs. Ho:YAG's 200 μm minimum), offering a better deflection range, improving accessibility, reducing impact on the operating channel, and ensuring adequate irrigation flow [6, 7]. The laser's electronic modulation capability offers unprecedented parameter flexibility, including pulse frequencies up to 2,200 Hz (44 × higher than Ho:YAG's 50 Hz limit), ultra-low pulse energies (0.005 J vs. 0.2 J minimum), and power outputs reaching 55 W (compared to Ho:YAG's typical 30 W maximum) [8, 9]. These features enhance surgical precision by improving perfusion, providing a clearer field of view, and allowing greater deflection angles for flexible scopes [5, 10]. The advantages of TFL make it a promising alternative for lithotripsy. Studies suggest TFL outperforms Ho:YAG lasers in operating time, complication rates, and stone-free rates (SFR) [3, 11,12,13,14].
The recent development of super-pulsed thulium fiber laser (SP TFL) technology represents a further advancement, combining high-frequency operation with ultra-short pulses to achieve more efficient fragmentation with minimal heat generation [15]. Preliminary evidence indicates that SP TFL achieves faster fragmentation and improved stone clearance compared to Ho:YAG lasers, without requiring higher energy inputs [10]. However, despite its promising advantages, SP-TFL was not approved for clinical use until 2017 by the Russian Ministry of Health, followed by FDA approval in 2019 and CE certification in 2020; this technology remains relatively uncommon in most urology departments around the world [9].
Due to its various advantageous characteristics, it may challenge Ho:YAG as the preferred laser for treatment; it could even represent a turning point in endourological procedures. However, given the novelty of SP TFL technology, it remains unclear whether TFL is superior to Ho:YAG in the treatment of patients with urolithiasis. The lack of robust comparative studies creates a critical knowledge gap regarding the optimal laser technology for endoscopic lithotripsy. A thorough evaluation of this issue holds significant clinical relevance. This review systematically evaluates the efficacy, safety, and clinical outcomes of SP-TFL vs. Ho:YAG laser lithotripsy, aiming to provide evidence-based insights to optimize urolithiasis management and improve patient care.
Materials and methods
This review was based on the recommendations of the Cochrane Collaboration for Systematic Reviews and reported in accordance with the PRISMA statement [16].
Eligibility criteria
According to PICOS [17], the inclusion criteria are listed as follows:
Population:
In‑vitro and preclinical studies: Synthetic stone models, ex vivo tissue models and animal models.
Clinical studies: Patients (adults/children) with urinary tract calculi (renal/ureteral stones) undergoing endoscopic lithotripsy procedures.
Interventions: SP TFL lithotripsy.
Comparator: Ho:YAG lithotripsy.
Outcomes:
In‑vitro and preclinical studies: Ablation rate, fragmentation, retropulsion, temperature and image distortion.
Clinical studies: Ablation speed and ablation efficiency, laser time and operative time, stone-free rate (SFR), retropulsion, visibility, and complication.
Exclusion criteria: Studies lacking direct comparisons between SP TFL and Ho:YAG. Case reports, editorials, letters, comments, conference abstracts. Nonrelevant studies after review of titles and abstracts.
Information sources and search strategy
For systematic review of currently available evidence on the Ho:YAG vs. SP TFL lithotripsy, a bibliographic search on PubMed, Embase, and Web of Science databases was conducted in September 2024. The systematic search strategy incorporated the following keywords, tailored to each database: “Super-Pulsed Thulium fiber laser”, “SP TFL”, “thulium”, “TEL”, “holmium:yttrium–aluminum-garnet”, “Holmium:YAG”, “Ho:YAG”, “kidney stone”, “urinary calculus”, “renal stone”, “lithotripsy”, “urolithiasis”. The following search strategy was used in each database: (("Super-Pulsed Thulium fiber laser" OR "SP TFL" OR "thulium fiber laser" OR "TFL") AND ("holmium:yttrium–aluminum-garnet" OR "holmium:YAG" OR "Ho:YAG") AND ("urolithiasis" OR "kidney stone" OR "ureteral stone" OR "urinary calculus" OR "renal stone" OR "lithotripsy")). A language restriction to English was applied, and studies published up to September 30, 2024, were included. Only original articles were considered eligible. Articles with irrelevant titles were eliminated in the first stage of the study selection process. In the second stage, abstracts and full texts of articles were examined to include those that satisfied the inclusion criteria. Use endnote X9 to arrange and assess headings and summaries, as well as to detect duplicate entries. Two evaluators (QG and JL) independently performed the search and cross-checked results to ensure accuracy. In instances where disagreements arose, a third evaluator (QZ) was consulted to achieve consensus, ensuring a balanced and unbiased selection process. A PRISMA 2020-compliant flow diagram has been included in Fig. 1, outlining the study identification, screening, eligibility assessment, and inclusion process.
Given the substantial heterogeneity among included studies—including variations in experimental designs, population characteristics, and outcome measures, a formal meta-analysis was not performed. Instead, a narrative review with qualitative synthesis was conducted. The methodological quality of the included studies was evaluated using the ROBINS-II for randomized controlled trials (RCTs) and the Modified Newcastle–Ottawa Scale (NOS) for observational studies [18, 19].
In‑vitro and preclinical studies (Table 1 )
Ablation rate
In vitro studies by Chew BH et al. found that the rate of the Ho:YAG laser to ablate a stone into fragments < 1 mm, when using both fragmentation and dusting settings, was lower than that of the SP TFL. The study highlighted that while both lasers are effective for stone fragmentation, SP TFL demonstrated superior efficiency in both fragmentation and dusting settings, achieving finer stone dust with a higher ablation rate [15]. Panthier et al. demonstrated that SP TFL achieves significantly higher ablation rates in dusting and fragmentation modes. For hard stones, SP TFL exhibits fourfold and twofold greater ablation rates in dusting(61.1 ± 7.88 vs. 16.26 ± 1.17 mm3/min) and fragmentation modes(66.96 ± 11.39 vs. 31.74 ± 4.60 mm3/min), respectively, while for soft stones, these rates are threefold (62.88 ± 9.25 vs. 22.99 ± 1.23 mm3/min) and twofold (66.57 ± 10.8 vs. 38.6 ± 5.34 mm3/min) higher than Ho:YAG under matched settings, highlighting the enhanced performance of SP TFL in both settings [20]. This enhanced performance is further validated in a porcine kidney model, where SP TFL produced shorter ablation times (9 min vs. 27 min, P < 0.001), lower energy expenditure (8 vs. 26 kJ, P < 0.001), and higher stone clearance (73% vs. 45%, P = 0.001) [21]. Similarly, several other studies also found that the ablation rate of SP TFL was higher than that of Ho:YAG laser under the same laser energy and repetition rate settings [22,23,24,25,26]. The superior stone ablation efficiency of SP TFL compared to Ho:YAG lasers can be attributed to its wavelength of 1940 nm, which is closer to the water absorption peak. This results in four- to fivefold greater water absorption compared to Ho:YAG lasers [27]. The higher water absorption reduces the tissue ablation threshold, as the water contained in the pores near the stone surface is essential for efficiently absorbing the laser energy [28]. Consequently, this enables the use of equivalent pulse energy for more effective stone ablation, making SP TFL a more efficient option for laser lithotripsy.
Fragmentation
The in vitro experiments conducted by Chew et al. showed that the SP TFL lithotripsy produces fewer particles > 2 mm in stone fragmentation compared to Ho:YAG laser (2.10 vs. 7.20 fragments, P < 0.001). Most stones treated with SP TFL turn into dust < 0.5 mm [15]. A study found that the SP TFL was more effective in producing smaller stone fragments compared to the Ho:YAG laser and Ho:YAG–MOSES laser systems. The SP TFL laser typically produces stone fragments smaller than 2 mm, whereas the largest stone fragments generated by Ho:YAG and Ho:YAG–MOSES were 4 mm and 3.5 mm, respectively [29]. After SP TFL lithotripsy, 77% of the remaining fragments were ≤ 1 mm, while 17% of the remaining fragments ≤ 1 mm after Ho:YAG treatment, which was reported in the study by Jiang et al. [21]. This suggests that SP TFL may offer advantages in reducing the size of stone fragments, potentially enhancing the ease of fragment removal and improving procedural efficiency.
Retropulsion
Retropulsion is a critical factor influencing the efficiency of lithotripsy. The difference in retropulsion force may have implications for the effectiveness and safety of lithotripsy procedures, with weaker retropulsion potentially minimizing stone migration during treatment. A study by Andreeva et al. demonstrated that when the pulse energy of the Ho:YAG laser was set at 0.2 J, stone displacement began. In contrast, with the SP TFL, stone displacement was observed at a pulse energy of 1 J with a 500 W peak power setting. Notably, for equal pulse energies, the retropulsion generated by the SP TFL was weaker than that of the Ho:YAG laser [22]. Ventimiglia et al. compared the stone displacement speed of SP TFL and Ho:YAG laser under different laser modalities. Their findings revealed that the average stone displacement for Ho:YAG in short pulse and long/Moses pulse modes (21–8.6 mm) was approximately 4 times and 2 times greater than that of SP TFL (4.5–4.9 mm), respectively. The average stone speed at the first moment of retropulsion was 4 and 1.8 times lower for the regular pulse mode of SP TFL (28 mm/s) than for Ho:YAG in short pulse and long/Moses pulse modes (141–47 mm/s), respectively. Retropulsion for SP TFL was lower than for Ho:YAG in different modes [23].
Temperature
In vitro and isolated studies
Andreeva V et al. monitored water temperature using four IT-23 thermocouples placed at the inlet, outlet, and two diagonally opposed locations adjacent to the stone at the base of the experimental tube. Irrigation flow was maintained at 24 mL/min (< 1 W) or 40 mL/min (≥ 16 W). No significant temperature differences were observed between SP-TFL and Ho:YAG lasers at equivalent power settings (8W: 4.9 ± 0.1 °C; 16W: 9.8 ± 0.1 °C; 40 W: 14.6 ± 0.1 °C) [22]. Taratkin M et al. employed T-type thermocouples to precisely monitor temperature variations during laser irradiation. The study was conducted in a thermally insulated test tube containing 37 ℃ warm water and an irrigated setup with flow rates set at 0, 10, and 35 mL/min, with a constant irradiation duration of 60 s. The results revealed that SP TFL induced a water temperature increase of 15.4 °C, compared to 14.9 °C for the Ho:YAG laser. The temperature rise of between the two lasers was similar. In addition, no statistically significant differences in temperature changes were observed between the two laser systems across the different irrigation flow rates [30].
In an ex vivo porcine kidney and ureter model, Molina WR et al. performed lithotripsy using both SP TFL and Ho:YAG lasers, with 5-s laser activation repeated 15 times for both dusting (SP TFL—0.1 J, 200 Hz, Short Pulse; Ho:YAG—0.3 J, 70 Hz, Long Pulse) and fragmentation settings (0.8 J, 8 Hz, Short Pulse for both). SPTFL generated higher temperatures in fragmentation mode (33.3 °C vs. 30.0 °C, P = 0.004), while Ho:YAG showed a trend toward higher temperatures in dusting mode (40.6 °C vs. 35.8 °C, P = 0.064). All temperatures remained below the 43 °C safety threshold, with no histological damage observed [31]. In a three-dimensional printed kidney–ureter model, Belle JD et al. demonstrated that under standardized irrigation (35 mL/min) with 60-s laser activation across four power settings (3.6, 10, 20, and 30 W), SP TFL consistently generated significantly higher average ureteral fluid temperatures than the Ho:YAG laser at all power levels (P < 0.001). While maximum temperatures remained below the critical 43 °C threshold for both lasers at 3.6,10 and 20 W, the SP TFL exceeded 43 °C at 30 W [32].
In vivo studies
In a porcine kidney model, Jiang P et al. conducted flexible ureteroscopic lithotripsy through a 14F ureteral access sheath, with both SP TFL (0.2 J, 80 Hz) and Ho:YAG laser (0.4 J, 40 Hz) operating in dusting mode. Continuous temperature monitoring using K-type thermocouples positioned in the intra-renal and renal pelvis revealed comparable thermal profiles between the two laser modalities during the procedure, with maximum recorded temperatures consistently remaining below 44 °C in both groups [21]. In another study, the researchers compared the effects of the two lasers on urothelium using three pig models. Each pig was treated with TFL on the left urinary tract and Ho:YAG laser on the right, and a total of 23 tests were performed. The temperature changes caused by the two lasers during treatment were similar, and no difference was found after anatomic pathology evaluation 3 weeks later. In the third pig, Bellini tubes were visible in both kidneys due to the intentional reduction of perfusion during temperature testing [25].
Previous studies have set the threshold for tissue damage at 43 ℃ [33, 34]. SP TFL’s higher thermal efficiency may elevate temperatures at high-power settings [32]. Clinicians should prioritize intermittent activation and adequate irrigation to keep temperatures below 43 °C and reduce the risk of tissue damage. In most preclinical models, both lasers maintained a safety threshold (< 43 °C). Future human trials under different surgical conditions are needed to verify the safety of both lasers.
Laser fiber fracture
Uzan et al. compared laser fiber fracture risks between 30W Ho:YAG and 50W SP TFL under varying fiber diameters, settings, and bending radii. SP TFL exhibited no fiber fractures across all conditions, demonstrating superior safety. In contrast, Ho:YAG showed significantly higher fracture rates, particularly with short pulses and high-energy settings [35]. Fiber fracture in Ho:YAG laser may affect the efficiency of surgery and increase costs. Intraoperative fiber replacement necessitated by fractures prolongs operative time. There is an increased risk of Ho:YAG laser rupture in anatomically challenging situations, such as renal pelvis stones that require active bending of fibers. In contrast, the fracture resistance of the SP TFL may have an advantage over the Ho:YAG laser in the treatment of renal pelvis stones.
Image distortion
Studies by Miller CS concluded that increased total power during flexible ureteroscopy causes image distortion to occur at greater distances from the tip of the ureteroscope. When comparing laser setups, the Ho:YAG laser produces image interference at greater distances than the SP TFL [36]. Image distortion in laser lithotripsy poses a significant clinical concern [37]. In laser lithotripsy, any visual distortion can compromise visibility and surgical precision, potentially prolonging operative time, increasing technical difficulty and the risk of accidental tissue damage [38]. Ultimately, this may adversely impact SFR and overall treatment efficacy.
Clinical studies (Table 2 )
Ablation speed and ablation efficiency
A prospective randomized clinical trial by Taratkin M et al. compared the clinical efficacy of SP TFL with Ho:YAG in retrograde intrarenal surgery (RIRS). The study found that SP TFL required less total energy for stone ablation than Ho:YAG (8.4 vs. 15.2 kJ, P = 0.021). Ablation speeds were similar between SP TFL (median 0.8 mm3/s) and Ho:YAG (0.6 mm3/s). However, the median ablation efficiency of SP TFL was lower than that of Ho:YAG laser (24.5 vs. 31.8 kJ/cm3, P = 0.046) [39].
For percutaneous nephrolithotomy (PCNL), Patil et al. reported comparable fragmentation speeds between SP TFL and Ho:YAG with MOSES technology (5.22 mm3/s vs. 4.64 mm3/s, P = 0.23), with no significant difference in total energy consumption (SP TFL: 16.30 ± 15.04 vs. Ho:YAG: 21.88 ± 23.32 kJ, P = 0.09). Both lasers exhibited similar efficacy in stone fragmentation and dusting across various stone densities [40]. Vergamini et al. 's study found no significant difference in ablation speed between SP TFL and Ho:YAG with MOSES technology (3.2 vs. 4.8 mm3/s, P > 0.05). However, SP TFL showed lower ablation efficiency (5.2 kJ/cm3) compared to the Ho:YAG (5.2 vs. 10.3 kJ/cm3, P < 0.001), and higher energy consumption (31.9 kJ vs. 11.8 kJ, P < 0.001) [41].
While these studies report various outcomes regarding ablation speed, energy consumption, and efficiency, there are notable discrepancies, primarily due to differences in laser parameters, surgical modalities and patient demographics. Future studies should aim for more homogeneous study designs, with standardized protocols and consistent reporting formats, to better elucidate the comparative effectiveness of SP TFL and Ho:YAG lasers in different clinical contexts.
Laser time and operative time
The SP TFL demonstrates advantages over the Ho:YAG laser in reducing operative and laser times across different lithotripsy procedures. Martov et al. observed that SP TFL was associated with shorter operative (24.7 ± 0.7 vs. 32.4 ± 0.7 min, P < 0.05) and laser times (8.4 ± 0.4 vs. 15.9 ± 0.5 min, P < 0.05) compared to Ho:YAG in ureterolithotripsy [10]. Similarly, Ulvik et al. reported a 14% shorter total operative time for SP TFL in ureterorenoscopic lithotripsy (49 vs. 57 min, P = 0.008), despite comparable laser times between the two lasers [3]. Ryan et al. further corroborated these findings, noting a 21% reduction in operative time with SP TFL (49.8 ± 29.1 vs. 62.8 ± 26.3 min, P = 0.021) [42]. Jaeger CD et al. found a longer laser time for SP TFL in pediatric ureteroscopy (11 vs. 2 min, P < 0.05), but no significant difference in total operative time (78 vs. 84 min, P > 0.05) [43]. In a comparative analysis of laser lithotripsy, SP TFL exhibited superior efficiency for 1–2 cm calculi, with operative times reduced by 18% compared to Ho:YAG (56.6 vs. 65.6 min, P = 0.04). However, this benefit did not extend to < 1 cm or > 2 cm stones [44].
For RIRS, Taratkin M et al. reported shorter laser times with SP TFL (9.2 vs. 14.1 min, P < 0.05); however, the reduction in operative time was not statistically significant (74.4 vs. 90.0 min, P = 0.054) [39]. While three studies found no difference in operative time between the two groups for RIRS [45,46,47].
Comparative studies reported inconsistent operative time for SP TFL in mini-PCNL. Mahajan AD et al. found that SP TFL was superior to the Ho:YAG laser in terms of laser time (11.19 vs. 20.45 min, P < 0.01) and operative time (55 vs. 68 min, P = 0.01) [11]. Patil et al. observed no difference in laser time but reported shorter operative times with SP TFL (23.68 vs. 41.86 min, P < 0.05) [40]. On the contrary, Vergamini et al. observed prolonged operative (91 vs. 67 min, P < 0.05) and laser times (17 vs. 7 min, P < 0.05) compared to Ho:YAG with MOSES in mini-PCNL [41]. Most studies support the advantages of SP TFL in shortening operative time. However, there is heterogeneity in the results, which may be related to differences in surgical procedures, surgical parameters, and patient populations. The current data are difficult to directly compare due to differences in surgical procedures and patient choices, and more standardized studies are needed.
SFR
Existing research indicated that the SFR of SP TFL was comparable to or even higher than that of the Ho:YAG laser. A prospective randomized clinical trial reported higher SFR for SP TFL compared to Ho:YAG (92% vs. 67%, P = 0.001) 3 months post-ureterorenoscopic lithotripsy, with a notable advantage for renal stones (86% vs. 49%, P = 0.001) [3]. Martov AG et al. found no stone residue in the SP TFL group and 5 patients with stone residue in the Ho:YAG laser group through non-contrast CT evaluation 1 month after ureteroscopy lithotripsy (0 vs. 6%) [10]. Jaeger CD et al. identified SFR as the absence of any stone fragments on the first imaging examination within 90 days after pediatric ureteroscopy. They observed that the SP TFL group had a higher SFR (70% vs. 59%, P < 0.05) [43]. Geavlete B et al. evaluated the SFR following RIRS via secondary ureteroscopy, revealing that 96.61% of patients in the SP TFL group achieved a stone-free status, compared to 86.63% in the Ho:YAG laser group [45]. Delbarre et al. found similar overall SFRs between the lasers; however, for stones measuring 1–2 cm, SP TFL achieved a higher SFR (81.6% vs. 62.5%, P = 0.04) [44]. Conversely, Haas CR et al., Taratkin et al., and Candela et al. reported no significant difference in SFR between SP TFL and Ho:YAG laser after RIRS [39, 46, 47].
For mini-PCNL, Mahajan AD et al. observed no statistical difference in SFR between SP TFL and Ho:YAG laser (94.9% vs. 90.9%, P = 0.498) [11]. Patil et al. found that the proportion of dusting (< 1 mm fragments) and SFR at 48 hours were similar between HPH-M and SP TFL [40]. Vergamini LB et al. also did not observe a difference in the SFR between the two groups, whether the SFR was defined as no fragments, no fragments > 2 mm, or no fragments > 4 mm (P = 0.52, 0.64 and 1.00, respectively) [41].
Retropulsion and visibility
The research results of Martov AG et al. showed that almost all (96%) cases in the SP TFL group experienced no observe retropulsion, while 31% of cases in the Ho:YAG laser group experienced retropulsion (P < 0.05). In 51% of cases in the Ho:YAG laser group and 4% of cases in the SP TFL group, mild retropulsion was observed (P < 0.05). Additionally, 0.18% of cases in the Ho:YAG laser group showed severe stone retropulsion, while no severe stone retropulsion was observed in the SP TFL group. SP TFL provided better endoscopic view quality, with 87% of patients in the SP TFL group having a clear endoscopic field, more than those in Ho:YAG laser group (87% vs. 64%, P < 0.05). The proportion of mild blurred visual field and severe blurred visual field in SP TFL group was lower than that in Ho:YAG laser group(12% vs. 30% and 1% vs. 6%, P < 0.05, respectively) [10].
Complication
In Ulvik et al.'s study, the proportion of intraoperative bleeding in the SP TFL group was lower than that in the Ho:YAG group (5% vs. 22%, P = 0.014) [3]. Martov AG et al. reported similar proportions of Clavien–Dindo I and II complications in the SP TFL group and the Ho:YAG laser group (10% vs. 13% and 1% vs. 5%, respectively) [10]. In Jaeger CD et al.'s retrospective study, no difference was found in the incidence of complications between SP TFL and Ho:YAG laser (25% vs. 22%, P > 0.05) [43]. Similarly, several studies have reported no difference in complication rates between the two groups [11, 39,40,41, 46, 47].
Discussion
The introduction of SP TFL provides a promising advancement in the field of urinary lithotripsy. Compared to the Ho:YAG, SP TFL offers several distinct advantages, including higher efficiency, a broader range of operational parameters, reduced retropulsion, enhanced visualization, and improved safety profiles. SP TFL, utilizing advanced pulse modulation technology, produces extremely high peak power, making it a more efficient option for stone fragmentation. This review aims to evaluate the current realities of SP TFL and Ho:YAG laser, with the goal of offering readers a balanced, evidence-based perspective.
Ablation efficiency
The results of in vitro studies indicated that SP TFL had higher ablation efficiency, particularly in dusting and fragmentation modes. For hard stones, SP TFL exhibits fourfold and twofold greater ablation rates in dusting and fragmentation modes, respectively, compared to Ho:YAG, while for soft stones, these rates are threefold and twofold higher under equivalent settings [15, 20]. This superiority is further corroborated by in vivo models, where SP TFL reduced ablation time by 67% (9 vs. 27 min, P < 0.001) and energy consumption by 69% (8 vs. 26 kJ, P < 0.001), alongside superior stone clearance (73% vs. 45%, P = 0.001) [21]. This is attributable to its 1940 nm wavelength, which exhibits a fourfold higher water absorption coefficient than the Ho:YAG laser. This unique optical property enables a "water-explosion" mechanism: water trapped within microcrystalline gaps, pores, and cracks inside and near the surface of the stone suddenly evaporates explosively, generating extremely high pressure in localized areas. These high pressures create mechanical stress waves within the stone, further contributing to its ablation, reducing the stone ablation threshold by a factor of four [5, 28]. In vitro studies by Hardy LA et al. showed that SP TFL had higher ablation efficiency than Ho:YAG laser under the action of this mechanism [48]. However, this mechanism has not been directly proven to impact Ho:YAG laser ablation [5]. In addition, SP TFL generated less retropulsion compared to Ho:YAG, which reduced the number of times the fiber moves toward the stone. This characteristic, combined with the water-explosion mechanism, allows SP TFL to achieve a comparable ablation rate while using less total ablation energy [39].
Retropulsion
SP TFL generates less retropulsion than Ho:YAG laser due to its lower peak power and smaller cavitation bubbles, which are four times smaller than those produced by Ho:YAG laser. Ventimiglia et al. quantified this disparity, revealing that Ho:YAG in short-pulse and long/Moses-pulse modes induced fourfold and twofold greater stone displacement distances (21–8.6 mm vs. 4.5–4.9 mm for SP TFL, P < 0.001) and 4–1.8 times higher initial retropulsion velocities (141–47 mm/s vs. 28 mm/s for SP TFL, P < 0.001) [23].SP TFL generates local pressures that are ten times lower than those of the Ho:YAG laser, contributing to reduced retropulsion [49]. The smaller retropulsion reduces the need to change the position of the optical fiber to adapt to the position of the stone. This leads to shorter operative times, improved efficiency in stone fragmentation, and a lower skill threshold for inexperienced surgeons, making SP TFL a more user-friendly option for lithotripsy.
Visibility
Due to the phenomenon of "snowstorm" in Ho:YAG laser lithotripsy, the clarity of the visual field during the operation is reduced. SP TFL has a better endoscopic field of view due to the smaller retropulsion and the consequent reduction in medium turbulence, resulting in less debris and dust particles being swept up, thus reducing the "snowstorm" effect [5]. A clear field of view reduces the number of interruptions in laser operations, shortens surgical time, and also lowers the risk of kidney and ureteral injuries.
Operative time
Multiple studies have investigated the impact of SP TFL and Ho:YAG laser on operative time and laser time. In ureteroscopy, SP TFL reduced total operative time by 14–21% and laser time by up to 47% [3, 10, 42]. In mini-PCNL, SP TFL exhibited conflicting outcomes: two studies reported shorter operative and laser times [11, 40], while another observed prolonged operative and laser times [41]. Most findings suggest that SP TFL offers advantages in reducing procedural duration, due to its smaller retropulsion, clearer field of view and higher ablation efficiency [3, 10, 11, 39, 40, 42]. Shorter operative time not only reduces surgeon fatigue but also decreases the risk of infection associated with prolonged operative time. However, it is worth noting that the study by Vergamini LB et al. found that the Ho:YAG laser had a shorter operative time in mini-PCNL [41]. Vergamini LB suggested that this may be due to differences in laser energy characteristics. The SP TFL absorbs four times more energy in water than the Ho:YAG laser and produces smaller pieces of stone. In PCNL surgery, the percutaneous sheath can remove the medium-sized stone fragments from the body without the need for SP TFL to break the stone into thinner fragments. The postoperative SFR showed no difference between the two groups, regardless of whether SFR was defined as no fragments, no fragments > 2 mm, or no fragments > 4 mm [41]. Two additional studies reported that SP TFL provided shorter operative times and similar SFR, compared to the Ho:YAG laser [11, 40]. While the overall evidence supports the efficiency of SP TFL, variations in surgical techniques, laser settings, and patient characteristics contribute to the observed inconsistencies across studies. Given the heterogeneity in current data, further standardized studies with uniform surgical protocols and laser parameters are necessary to better define the comparative advantages of SP TFL and Ho:YAG laser in different clinical situations.
SFR
Existing research indicated that the SFR of SP TFL was comparable to or even higher than that of the Ho:YAG laser [10, 39,40,41, 43]. In ureterorenoscopic lithotripsy, a prospective randomized trial reported a significantly higher SFR for SP TFL at 3 months postoperatively (92% vs. 67%, P = 0.001), with pronounced efficacy for renal stones (86% vs. 49%, P = 0.001) [3]. Martov et al. further supported this advantage, showing no residual stones in the SP TFL group vs. 6% in the Ho:YAG group at 1 month (P < 0.05) [10]. Pediatric ureteroscopy data also favored SP TFL, with a 70% SFR compared to 59% for Ho:YAG (P < 0.05) [34]. Residual stone fragments after treatment may become the focus of stone regeneration, recurrence, urinary tract infection and ureteral obstruction [50]. This may become the reason for re-treatment, affecting the health of patients and increasing the burden on patients. Traditionally, non-obstructive stone fragments ≤ 4 mm in diameter have been described as clinically irrelevant residual fragments, but this concept has been questioned, because up to 26% of patients with residual stones develop symptoms and require intervention within 2 years of treatment [50, 51]. Studies have shown that stone fragments ≤ 2 mm were less likely to grow over time, more likely to pass spontaneously, and had lower rates of regrowth, clinical events, and reintervention [52, 53]. Therefore, stone-free status or smaller stone fragments are of great significance for the prognosis of patients. The results of the in vitro study showed that SP TFL produced a higher proportion of stone fragments < 2 mm and a larger quantity of dust [15, 21, 29]. The clinical results showed that the SFR of SP TFL was better than that of the Ho:YAG laser [10, 43]. This is related to the higher water absorption of SP TFL and the resulting micro-explosive water vapor expansion mechanism, as well as the clearly defined and precise emission of laser energy [5, 48]. With its wide-ranging and flexible parameters, the SP TFL is also capable of high-frequency dusting, pop-corning, pop-dusting, corn-dusting, and micro-dusting. SP TFL can produce smaller stone fragments, reducing the chance of residual debris after stone surgery. This could ultimately reduce repeat surgery and related clinical events for patients. Notably, there was a lack of significant SFR differences between the two in some studies, highlighting the influence of factors, such as surgical technique, stone characteristics, and follow-up time. Further large-scale, multicenter trials with standardized definitions of SFR are needed to clarify the clinical benefits of SP TFL over Ho:YAG laser across different lithotripsy modalities.
Thermal damage
The thermal damage resulting from increased temperatures during laser lithotripsy is a recognized issue. Due to the higher water absorption rate of SP TFL, this may lead to temperature increases during stone fragmentation, potentially causing thermal damage [5]. In vitro studies suggest that both SP TFL and Ho:YAG laser are safe under typical lithotripsy conditions with appropriate irrigation [21, 22, 30, 31]. However, SP TFL demonstrated a trend of generating slightly higher temperatures in fragmentation mode and at higher power settings [31, 32], highlighting potential thermal risks at high-power settings. According to Sapareto et al., 43 °C was the threshold temperature for inducing tissue damage. At this temperature, a duration of 120 min is required to trigger thermal injury, highlighting the close correlation between cellular damage and exposure time [33]. Clinically, these findings underscore the importance of optimizing irrigation flow rates, as increased irrigation can effectively mitigate temperature elevation [54], particularly at higher power settings, as increased power can lead to significant rises in intra-ureteral temperature, elevating the risk of thermal injury to the ureter [26]. While SP TFL's higher efficiency in stone fragmentation may reduce total energy application and procedural duration, minimizing prolonged laser activation and ensuring adequate irrigation are essential to prevent thermal injury. At present, there is a lack of clinical studies on the thermal damage caused by two types of lasers. Further research should investigate potential thermal injury manifestations, such as ureteral stricture formation, especially in cases where high-power settings or prolonged activation are used. In addition, the interaction of SP TFL’s higher water absorption with different irrigation strategies and laser settings requires further refinement to optimize clinical protocols.
Laser fiber fracture and image distortion
Uzan A et al. found that SP TFL did not experience fiber breakage regardless of the laser setting, fiber diameter, and bending radius. In contrast, the Ho:YAG laser demonstrated a higher risk of fiber breakage compared to SP TFL [35]. In addition, Miller CS et al. reported that increased laser power led to image distortion at a distance from the ureteroscope tip. At the same laser setup, Ho:YAG produced image interference at a greater distance than SP TFL [36]. The reduced risk of fiber breakage across various laser settings and fiber configurations suggests greater durability and operational reliability during lithotripsy procedures, potentially lowering the need for frequent fiber replacement and reducing procedural interruptions. Furthermore, the superior endoscopic visibility associated with SP TFL, particularly under high-power settings, enhances intraoperative precision and safety. These advantages may contribute to improved procedural efficiency and surgical outcomes, supporting that SP TFL may offer a safer option for lithotripsy.
Implications for clinical practice
SP TFL demonstrates excellent ablation efficiency, shortening operative time while producing finer fragments. This translates into improved surgical efficiency, reduced anesthesia exposure, a lower risk of infections associated with prolonged surgery, and enhanced operating room turnover. A comparable or superior SFR means lower residual stone volume, reducing the risk of infection, obstruction, and the need for secondary procedures. The SP TFL has a smaller retropulsion force and better visualization, which improves surgical precision and reduces the need for instrument repositioning and interruptions to restore the visual field. The breakage resistance of the SP TFL fiber means it can meet the bending requirements in anatomically challenging cases, such as renal calyx stones. In addition, SP TFL shares a similar safety profile with Ho:YAG lasers, and in vitro experiments show similar temperature variations at the same output power. Therefore, strategies used for Ho:YAG lasers, such as using cold irrigation fluids and intermittent laser firing to minimize thermal effects, are also applicable to SP TFL, helping to prevent unnecessary damage to surrounding tissues. Overall, SP TFL represents a promising advancement in laser lithotripsy, offering improvements in visualization, handling, and procedural efficiency without compromising safety or efficacy. However, inconsistencies still exist across some studies, likely due to variations in surgical technique, patient characteristics, and energy settings.
Limitation
The limitations of this review should be acknowledged. The number of studies on this new technology is limited and the number of studies on which conclusions can be drawn is small. Many studies have small sample sizes and are conducted in single centers, which may introduce selection and performance bias. Furthermore, there was significant heterogeneity in surgical techniques, laser settings and outcome definitions, making direct comparison across studies challenging. To better define the clinical advantages of SP TFL over Ho:YAG laser, future research should prioritize well-designed randomized controlled trials with larger, multicenter cohorts. In addition, studies focusing on long-term outcomes such as complication rates (e.g., ureteral strictures), cost-effectiveness, and procedural learning curves are warranted to guide more comprehensive clinical decision-making. Finally, although the parameter settings and effects of lithotripsy mode have been reported in the literature, the ideal SP TFL lithotripsy parameters have not been determined, and more high-quality clinical studies are still needed to evaluate the efficiency and safety of SP TFL lithotripsy at different parameter settings.
Conclusion
The emergence of SP TFL offers a promising new option for the minimally invasive treatment of urinary calculi. SP TFL has several advantages over the Ho:YAG laser, including enhanced lithotripsy efficiency, more versatile and comprehensive parameter settings, significantly reduced retropulsion, and a higher degree of miniaturization. These benefits contribute to more effective and controlled treatments, especially in cases, such as ureterorenoscopic lithotripsy and RIRS, where SP TFL has shown a shorter operative time and a higher stone-free rate compared to Ho:YAG laser. Clinicians need to understand the technical characteristics and settings of SP TFL under different conditions to apply the most appropriate techniques for various procedural approaches. Despite its potential, SP TFL has a relatively short clinical application history, necessitating further research to fully elucidate its long-term advantages, clinical significance, and possible limitations. Future studies should focus on in vivo evaluations, randomized clinical trials, and cost-effectiveness analyses to provide comprehensive insights into the utility of SP TFL in urological practice.
Availability of data and materials
The data sets used or analyzed during the current study are available from the corresponding author on reasonable request. No datasets were generated or analysed during the current study.
Abbreviations
- SP TFL:
-
SuperPulsed thulium fiber laser
- Ho:YAG:
-
Holmium:yttrium–aluminum–garnet
- TEL:
-
Thulium fiber laser
- RIRS:
-
Retrograde intrarenal surgery
- PCNL:
-
Percutaneous nephrolithotomy
References
Elhilali MM, Badaan S, Ibrahim A, Andonian S. Use of the moses technology to improve holmium laser lithotripsy outcomes: a preclinical study. J Endourol. 2017;31(6):598–604.
Yu S, Liu L, Li Y, Zhou L, Chen J, Li H, et al. Flexible ureteroscopic treatment of kidney stones: how do the new laser systems change our concepts? Asian J Urol. 2024;11(2):156–68.
Ulvik O, Æsøy MS, Juliebo-Jones P, Gjengsto P, Beisland C. Thulium fibre laser versus Holmium:YAG for ureteroscopic lithotripsy: outcomes from a prospective randomised clinical trial. Eur Urol. 2022;82(1):73–9.
Traxer O, Keller EX. Thulium fiber laser: the new player for kidney stone treatment? A comparison with Holmium:YAG laser. World J Urol. 2020;38(8):1883–94.
Kronenberg P, Traxer O. The laser of the future: reality and expectations about the new thulium fiber laser-a systematic review. Transl Androl Urol. 2019;8(Suppl 4):S398–417.
Enikeev D, Herrmann TRW, Taratkin M, Azilgareeva C, Borodina A, Traxer O. Thulium fiber laser in endourology: current clinical evidence. Curr Opin Urol. 2023;33(2):95–107.
Taratkin M, Azilgareeva C, Korolev D, Barghouthy Y, Tsarichenko D, Akopyan G, et al. Prospective single-center study of superpulsed thulium fiber laser in retrograde intrarenal surgery: initial clinical data. Urol Int. 2022;106(4):404–10.
Almasoud NA, Safar O, Elatreisy A, Alshahrani ST, Libdah SB, Alkhaldi SM, et al. Super pulsed thulium fiber laser outcomes in retrograde intrarenal surgery for ureteral and renal stones: a systematic review and meta-analysis. BMC Urol. 2023;23(1):179.
Kronenberg P, Hameed BMZ, Somani B. Outcomes of thulium fibre laser for treatment of urinary tract stones: results of a systematic review. Curr Opin Urol. 2021;31(2):80–6.
Martov AG, Ergakov DV, Guseynov M, Andronov AS, Plekhanova OA. Clinical comparison of super pulse thulium fiber laser and high-power holmium laser for ureteral stone management. J Endourol. 2021;35(6):795–800.
Mahajan AD, Mahajan SA. Thulium fiber laser versus holmium. Indian J Urol. 2022;38(1):42–7.
Uleri A, Farré A, Izquierdo P, Angerri O, Kanashiro A, Balaña J, et al. Thulium fiber laser versus holmium: yttrium aluminum garnet for lithotripsy: a systematic review and meta-analysis. Eur Urol. 2024;85(6):529–40.
Tang X, Wu S, Li Z, Wang D, Lei C, Liu T, et al. Comparison of thulium fiber laser versus holmium laser in ureteroscopic lithotripsy: a meta-analysis and systematic review. BMC Urol. 2024;24(1):44.
Chua ME, Bobrowski A, Ahmad I, Kim JK, Silangcruz JM, Rickard M, et al. Thulium fibre laser vs holmium: yttrium-aluminium-garnet laser lithotripsy for urolithiasis: meta-analysis of clinical studies. BJU Int. 2023;131(4):383–94.
Chew BH, Koo KC, Halawani A, Lundeen CJ, Knudsen BE, Molina WR. Comparing dusting and fragmenting efficiency using the new SuperPulsed thulium fiber laser versus a 120 W Holmium:YAG laser. Investig Clin Urol. 2023;64(3):265–71.
Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372: n71.
Cumpston M, Li T, Page MJ, Chandler J, Welch VA, Higgins JP et al. Updated guidance for trusted systematic reviews: a new edition of the Cochrane Handbook for Systematic Reviews of Interventions. Cochrane Database Syst Rev 2019; 10(10):ED000142.
Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, et al. The cochrane collaboration’s tool for assessing risk of bias in randomised trials. BMJ. 2011;343: d5928.
Stang A. Critical evaluation of the Newcastle-Ottawa scale for the assessment of the quality of nonrandomized studies in meta-analyses. Eur J Epidemiol. 2010;25(9):603–5.
Panthier F, Doizi S, Lapouge P, Chaussain C, Kogane N, Berthe L, et al. Comparison of the ablation rates, fissures and fragments produced with 150 microm and 272 microm laser fibers with superpulsed thulium fiber laser: an in vitro study. World J Urol. 2021;39(6):1683–91.
Jiang P, Okhunov Z, Afyouni AS, Ali S, Hosseini Sharifi SH, Bhatt R, et al. Comparison of superpulse thulium fiber laser vs holmium laser for ablation of renal calculi in an in vivo porcine model. J Endourol. 2023;37(3):335–40.
Andreeva V, Vinarov A, Yaroslavsky I, Kovalenko A, Vybornov A, Rapoport L, et al. Preclinical comparison of superpulse thulium fiber laser and a holmium:YAG laser for lithotripsy. World J Urol. 2020;38(2):497–503.
Ventimiglia E, Doizi S, Kovalenko A, Andreeva V, Traxer O. Effect of temporal pulse shape on urinary stone phantom retropulsion rate and ablation efficiency using holmium:YAG and super-pulse thulium fibre lasers. BJU Int. 2020;126(1):159–67.
Kutchukian S, Doizi S, Lapouge P, Germain T, Dragos L, Berthe L, et al. Ablation rates with Holmium:YAG and Thulium Fiber Laser: Influence of the stone phantom homogeneity. An in vitro study. Prog Urol. 2023;33(8–9):456–62.
Sierra A, Corrales M, Somani B, Traxer O. Laser efficiency and laser safety: holmium YAG vs. thulium fiber laser. J Clin Med. 2022;12(1):149.
Sierra A, Corrales M, Kolvatzis M, Panthier F, Pinero A, Traxer O. Thermal injury and laser efficiency with holmium YAG and thulium fiber laser-an in vitro study. J Endourol. 2022;36(12):1599–606.
Fried NM, Irby PB. Advances in laser technology and fibre-optic delivery systems in lithotripsy. Nat Rev Urol. 2018;15(9):563–73.
Fried NM. Recent advances in infrared laser lithotripsy [Invited]. Biomed Opt Express. 2018;9(9):4552–68.
Jiang P, Peta A, Brevik A, Arada RB, Ayad M, Afyouni AS, et al. Ex vivo renal stone dusting: impact of laser modality, ureteral access sheath, and suction on total stone clearance. J Endourol. 2022;36(4):499–507.
Taratkin M, Laukhtina E, Singla N, Kozlov V, Abdusalamov A, Ali S, et al. Temperature changes during laser lithotripsy with Ho:YAG laser and novel Tm-fiber laser: a comparative in-vitro study. World J Urol. 2020;38(12):3261–6.
Molina WR, Carrera RV, Chew BH, Knudsen BE. Temperature rise during ureteral laser lithotripsy: comparison of super pulse thulium fiber laser (SPTF) vs high power 120 W holmium-YAG laser (Ho:YAG). World J Urol. 2021;39(10):3951–6.
Belle JD, Chen R, Srikureja N, Amasyali AS, Keheila M, Baldwin DD. Does the novel thulium fiber laser have a higher risk of urothelial thermal injury than the conventional holmium laser in an in vitro study? J Endourol. 2022;36(9):1249–54.
Sapareto SA, Dewey WC. Thermal dose determination in cancer therapy. Int J Radiat Oncol Biol Phys. 1984;10(6):787–800.
Dewhirst MW, Viglianti BL, Lora-Michiels M, Hanson M, Hoopes PJ. Basic principles of thermal dosimetry and thermal thresholds for tissue damage from hyperthermia. Int J Hyperthermia. 2003;19(3):267–94.
Uzan A, Chiron P, Panthier F, Haddad M, Berthe L, Traxer O, et al. Comparison of Holmium:YAG and thulium fiber lasers on the risk of laser fiber fracture. J Clin Med. 2021;10(13):2960.
Miller CS, Whiles BB, Ito WE, Machen E, Thompson JA, Duchene DA, et al. Image distortion during flexible ureteroscopy: a laboratory model comparing super pulsed thulium fiber laser vs high-power Ho:YAG laser. J Endourol. 2023;37(1):99–104.
Rajamahanty S, Grasso M. Flexible ureteroscopy update: indications, instrumentation and technical advances. Indian J Urol. 2008;24(4):532–7.
Humphreys MR, Miller NL, Williams JC, Evan AP, Munch LC, Lingeman JE. A new world revealed: early experience with digital ureteroscopy. J Urol. 2008;179(3):970–5.
Taratkin M, Azilgareeva C, Petov V, Morozov A, Ali S, Babaevskaya D, et al. Thulium fiber laser vs Ho:YAG in RIRS: a prospective randomized clinical trial assessing the efficacy of lasers and different fiber diameters (150 microm and 200 microm). World J Urol. 2023;41(12):3705–11.
Patil A, Reddy N, Shah D, Singh A, Ganpule A, Sabnis R, et al. High-power holmium with MOSES technology or Thulium fiber laser in miniperc with suction: a new curiosity. J Endourol. 2022;36(10):1348–54.
Vergamini LB, Ito W, Choi BN, Du HE, Sardiu ME, Neff D, et al. Holmium:yttrium-aluminium-garnet laser with MOSES technology is more efficient than thulium fibre laser in supine mini-percutaneous nephrolithotomy. BJU Int. 2024;134(2):276–82.
Ryan JR, Nguyen MH, Linscott JA, Nowicki SW, James E, Jumper BM, et al. Ureteroscopy with thulium fiber laser lithotripsy results in shorter operating times and large cost savings. World J Urol. 2022;40(8):2077–82.
Jaeger CD, Nelson CP, Cilento BG, Logvinenko T, Kurtz MP. Comparing pediatric ureteroscopy outcomes with superpulsed thulium fiber laser and low-power Holmium:YAG Laser. J Urol. 2022;208(2):426–33.
Delbarre B, Baowaidan F, Culty T, Khelfat L, Brassier M, Ferragu M, et al. Prospective comparison of thulium and holmium laser lithotripsy for the treatment of upper urinary tract lithiasis. Eur Urol Open Sci. 2023;51:7–12.
Geavlete B, Mares C, Multescu R, Georgescu D, Geavlete P. Second-look flexible ureteroscopy after RIRS–Holmium Moses versus TFL (Soltive). J Med Life. 2022;15(10):1218–23.
Haas CR, Knoedler MA, Li S, Gralnek DR, Best SL, Penniston KL, et al. Pulse-modulated Holmium:YAG laser vs the thulium fiber laser for renal and ureteral stones: a single-center prospective randomized clinical trial. J Urol. 2023;209(2):374–83.
Candela L, Solano C, Castellani D, Teoh JY, Tanidir Y, Fong KY, et al. Comparing outcomes of thulium fiber laser versus high-power Holmium:YAG laser lithotripsy in pediatric patients managed with RIRS for kidney stones A multicenter retrospective study. Minerva Pediatr. 2023. https://doiorg.publicaciones.saludcastillayleon.es/10.23736/s2724-5276.23.07392-5.
Hardy LA, Vinnichenko V, Fried NM. High power holmium:YAG versus thulium fiber laser treatment of kidney stones in dusting mode: ablation rate and fragment size studies. Lasers Surg Med. 2019;51(6):522–30.
Hardy LA, Kennedy JD, Wilson CR, Irby PB, Fried NM. Analysis of thulium fiber laser induced bubble dynamics for ablation of kidney stones. J Biophotonics. 2017;10(10):1240–9.
Chew BH, Brotherhood HL, Sur RL, Wang AQ, Knudsen BE, Yong C, et al. Natural history, complications and re-intervention rates of asymptomatic residual stone fragments after ureteroscopy: a report from the EDGE research consortium. J Urol. 2016;195(41):982–6.
Suarez-Ibarrola R, Hein S, Miernik A. Residual stone fragments. Curr Opin Urol. 2019;29(2):129–34.
Wong VKF, Que J, Kong EK, Abedi G, Nimmagadda N, Emmott AS, et al. The fate of residual fragments after percutaneous nephrolithotomy: results from the endourologic disease group for excellence research consortium. J Endourol. 2023;37(6):617–22.
Panthier F, Kwok J-L, Tzou DT, Monga M, Traxer O, Keller EX. What is the definition of stone dust and how does it compare with clinically insignificant residual fragments? A comprehensive review. World J Urol. 2024;42(1):292.
Molina WR, Silva IN, Donalisio da Silva R, Gustafson D, Sehrt D, Kim FJ. Influence of saline on temperature profile of laser lithotripsy activation. J Endourol. 2015;29(2):235–9.
Acknowledgements
Not applicable.
Funding
This work was funded by the Guangxi Key Research and Development Program (No. Guike AB23075188) and Central-led Local Science and Technology Development Fund (No. Guike ZY24212064).
Author information
Authors and Affiliations
Contributions
Qinghua Gan, Jun Li and Gaoqiang Zhai wrote the main manuscript, Qinghua Gan, Gaoqiang Zhai, Bangfeng Liu, Jun Li, Yan Qin, Shuting Tan collected the literature and prepared the tables. Qinghua Gan, Jun Li, Wei Wang and Qinsong Zeng modified and revised the manuscript; Wei Wang and Qinsong Zeng supervised in the design of the study and finalized the manuscript. All authors have contributed to the manuscript and approved the submitted version.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Gan, Q., Li, J., Zhai, G. et al. Expanding urolithiasis treatment: comparison of super pulsed thulium laser and holmium:YAG laser for ureteral stone management. Eur J Med Res 30, 379 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40001-025-02591-7
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40001-025-02591-7